Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
New post-imaging software provides fast and accurate volume data from CTA surveillance after endovascular aneurysm
Kay K Yeung1, Maarten J van der Laan, Jan J Wever
1Department of Vascular Surgery, University Medical Center Utrecht, The Netherlands.
Purpose:
To quantify intra- and interobserver variabilities when measuring total aneurysm volume after endovascular aneurysm repair using the Vitrea 2 System and to compare it in terms of accuracy and processing time with the gold standard methods using the Easy Vision workstation.
Methods:
Total aneurysm volumes from 30 postendograft CTA datasets were randomly selected from a database consisting of approximately 400 CTA datasets recorded in 89 patients. The intra- and interobserver variabilities were measured on the Vitrea workstation by 2 investigators. The intermodality variability was calculated for the same measurements using the Easy Vision workstation. The differences of each pair of measurements were plotted against their mean, and the repeatability coefficient (RC) was calculated. The mean differences were also expressed as a percentage of the first measurements.
Results:
The intraobserver mean difference was 1.6 mL (1.4%) with an RC of 10.8 mL (10.1%) and the interobserver mean difference was -1.4 mL (-1.4%) with an RC of 11.7 mL (10.2%). The intermodality mean difference was 1.8 mL (2.0%) with an RC of 15.8 mL (11.1%). The Vitrea workstation required a median of 8 minutes (interquartile range 7-10) for 1 observer and 6 minutes (interquartile range 5-8) for the other to perform a complete volume segmentation of each patient dataset compared to an estimated average of 30 minutes using the Easy Vision workstation.
Conclusions:
The Vitrea workstation provides fast and accurate volume data from spiral CTA follow-up of endovascular aneurysm repair. This software may enhance the acceptability of volume surveillance in daily practice.