Are new-generation flexible ureteroscopes better than their predecessors?
1Division of Urology, Department of Surgery, Taichung Veterans General Hospital, Taiwan. chiu3778@hotmail.com
BJU International
|December 18, 2003
Summary
New flexible ureteroscopes offer significantly improved tip deflection for better diagnosis and treatment of lower pole renal pathology. However, these advancements come with reduced flow rates and increased costs.
Area of Science:
- Urology
- Medical Devices
- Endoscopy
Background:
- Flexible ureteroscopes are crucial for diagnosing and treating lower pole renal pathology.
- Tip deflection limitations can hinder effective diagnosis and treatment.
- Previous ureteroscope models faced challenges with maintaining deflection when instruments were inserted.
Purpose of the Study:
- To compare the tip deflection and performance of two new-generation flexible ureteroscopes against their predecessors.
- To evaluate if design advancements have improved the handling of lower pole renal pathology.
- To assess the impact of endoscopic tools on the tip deflection and flow rates of new ureteroscopes.
Main Methods:
- Two new flexible ureteroscopes (ACMI DUR-8 Elite, Storz 11278AU) were compared with their older models (ACMI DUR-8, Storz 11274AAU).
- Active tip deflection and irrigation flow rates were measured with and without endoscopic tools.
- Specifications, purchase prices, and repair costs were obtained; field of view and image size were also compared.
Main Results:
- The ACMI DUR-8 Elite and Storz 11278AU showed significant improvements in tip deflection (79 and 144 degrees, respectively).
- Newer models experienced less loss of tip deflection with endoscopic tools inserted (0.7% and 2.8% loss vs. 9.6% and 5.0%).
- Flow rates were decreased in the new ureteroscope models.
Conclusions:
- New flexible ureteroscopes demonstrate superior active tip deflection compared to older models, even with instruments inserted.
- Improved flexibility comes at the cost of reduced flow rates.
- Higher purchase costs are associated with the new ureteroscope designs.