Related Experiment Video
Updated: Aug 29, 2026

Controlled Synthesis and Fluorescence Tracking of Highly Uniform Poly(N-isopropylacrylamide) Microgels
Published on: September 8, 2016
Scatter-to-primary based scatter fractions for transmission-dependent convolution subtraction of SPECT images
Anne Larsson1, Lennart Johansson
1Department of Radiation Sciences, Radiation Physics, Umeå University, SE-901 87 Umeå, Sweden. anne.larsson@vll.se
Abstract:
In single photon emission computed tomography (SPECT), transmission-dependent convolution subtraction has been shown to be useful when correcting for scattered events. The method is based on convolution subtraction, but includes a matrix of scatter fractions instead of a global scatter fraction. The method can be extended to iteratively improve the scatter estimate, but in this note we show that this requires a modification of the theory to use scatter-to-total scatter fractions for the first iteration only and scatter-to-primary fractions thereafter. To demonstrate this, scatter correction is performed on a Monte Carlo simulated image of a point source of activity in water. The modification of the theory is compared to corrections where the scatter fractions are based on the scatter-to-total ratio, using one and ten iterations. The resulting ratios of subtracted to original counts are compared to the true scatter-to-total ratio of the simulation and the most accurate result is found for our modification of the theory.