Current limitations of volumetric capnography in surfactant-depleted small lungs
Hans Proquitté1, Susann Krause, Mario Rüdiger
1Clinic of Neonatology (Charité Campus Mitte), Humboldt-University, Berlin, Germany.
Objective:
To investigate the suitability of volumetric capnography for assessing alveolar gas exchange in very small, surfactant-depleted lungs.
Design:
Prospective animal trial.
Settings:
Animal laboratory in a university setting.
Subjects:
Twenty-one ventilated newborn piglets (age <12 hrs; median weight, 890 g; range, 560-1435 g).
Interventions:
Bronchoalveolar lavage with instillation of 30 mL/kg normal saline. Ventilatory, circulatory, and lung mechanic variables were measured before and 0, 30, and 60 mins after bronchoalveolar lavage.
Measurements And Main Results:
The alveolar deadspace fraction calculated by the Bohr and the Bohr/Enghoff equations increased three-fold (p<.001) after bronchoalveolar lavage in capnograms with distinct alveolar plateau, whereas in capnograms without alveolar plateau no statistical significant difference was seen. The main problem of capnography in small and especially stiff lungs was the high number of discarded records exclusively caused by a missing alveolar plateau. Rates of discarded records of capnography were 9.5% before lavage and increased (p<.01) to 52.4%, 47.6%,42.8% after bronchoalveolar lavage (0, 30, and 60 mins). With decreasing exhalation time, the number of discarded records increased significantly. No plateau was seen in >75% of recorded files with exhalation times <200 msecs. The effect of bronchoalveolar lavage on all variables measured was quite different, with the highest impact on required ventilatory settings, calculated oxygenation variables, and compliance. The effect of bronchoalveolar lavage on arterio-alveolar CO2 difference, CO2 production, and alveolar deadspace was much lower and statistically significant only in capnograms with alveolar plateau.
Conclusions:
Volumetric capnography is a useful tool to detect impaired alveolar gas exchange in surfactant-depleted small lungs. However, the method failed if there was no alveolar plateau in the volumetric capnogram especially in stiff lungs with short exhalation times.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Breathing
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Volumes and Capacities
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Acute Respiratory Failure-III