Clinical decision-making in hypertension using an automated (BpTRU) measurement device
1Mayo Clinic and Mayo Foundation, Rochester, MN 55905, USA. graves.john@mayo.edu
Abstract:
Mercury sphygmomanometers are being removed from clinical practice in the United States due to environmental concerns about mercury toxicity. Accurate blood pressure measurement is central to high-quality hypertension management. In this study of 106 patients, the BpTRU(TM) device was compared to nurse blood pressure measurements that complied with all the JNC VII/American Heart Association guidelines in evaluation of a random casual blood pressure. The intermethod difference in systolic blood pressure was +1.8+/-5.1 mmHg, and for diastolic blood pressure it was 4.8+/-5.1 mmHg (both P<0.001). For the primary study end point of clinical decision-making, there was 92% (97/106) agreement between the hypertension nurse specialist and the BpTRU (kappa 0.8280, 95% confidence interval, 0.721-0.9350). The oscillometric blood pressure measurement with the BpTRU is recommended as a replacement for poorly performed auscultatory blood pressure measurement in clinical practice.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Hypertension III: Clinical Manifestations and Diagnostic Studies