Comparison of traditional and plethysmographic methods for measuring pulsus paradoxus
Jeff A Clark1, Mary Lieh-Lai, Ron Thomas
1Critical Care Medicine, Department of Pediatrics, Children's Hospital of Michigan, Wayne State University School of Medicine, Detroit, 48201-2196, USA. jclark@crcm.med.wayne.edu
Background:
In the evaluation of patients with acute asthma, pulsus paradoxus (PP) is an objective and noninvasive indicator of the severity of airway obstruction. However, in children PP may be difficult or impossible to measure. Indwelling arterial catheters facilitate the measurement of PP, but they are invasive and generally reserved for critically ill patients.
Objective:
To determine the utility of the plethysmographic waveform (PPpleth) of the pulse oximeter in measuring PP.
Methods:
Patients from the pediatric intensive care unit, emergency department, and inpatient wards of a tertiary care pediatric hospital were eligible for the study. A total of 36 patients (mean age [SD], 11.2 [4.7] years) were enrolled in the study. Pulsus paradoxus was measured using the traditional auscultatory (PPausc) method with a sphygmomanometer. Pulsus paradoxus was then measured using a blood pressure cuff observing for the disappearance and reappearance of the (PPpleth) on the pulse oximeter. Mean difference and 95% confidence intervals were calculated for each method. The 2 methods were also analyzed for correlation and agreement using the Pearson product moment correlation and a Bland and Altman plot.
Results:
Patients with status asthmaticus had higher PPausc and PPpleth readings compared with nonasthmatic patients. Pulsus paradoxus measured by plethysmography in patients with and without asthma was similar to PPausc readings (mean difference, 0.6 mm Hg; 95% confidence interval, -0.6 to 2.1 mm Hg). Individual PPpleth readings showed significant correlation and agreement with PPausc readings in patients both with and without asthma.
Conclusion:
Measurement of PP using the pulse oximeter-pulse plethysmographic waveform offers a simple and noninvasive method for evaluating patients with airway obstruction.
Related Concept Videos
Measurement of Blood Pressure
Pulse
The pulse serves as a clinical indicator...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Special considerations while measuring pulse