Dosimetric IMRT verification with a flat-panel EPID.

B Warkentin1, S Steciw, S Rathee

  • 1Department of Medical Physics, Cross Cancer Institute, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada.

Medical Physics
|January 10, 2004
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Retraction: Minimal skin dose increase in longitudinal rotating biplanar linac-MR systems: examination of radiation energy and flattening filter design (2016<i>Phys. Med. Biol.</i>61 3527).

Physics in medicine and biology·2025
Same author

Seed size dimorphism in Hyptis suaveolens aids in differentiation of the germination niche.

Plant biology (Stuttgart, Germany)·2022
Same author

Feasibility of energy adaptive angular meshing for perpendicular and parallel magnetic fields in a grid based Boltzmann solver.

Biomedical physics & engineering express·2021
Same author

A novel transport sweep architecture for efficient deterministic patient dose calculations in MRI-guided radiotherapy.

Physics in medicine and biology·2019
Same author

Proton beam behavior in a parallel configured MRI-proton therapy hybrid: Effects of time-varying gradient magnetic fields.

Medical physics·2018
Same author

On the direct acquisition of beam's-eye-view images in MRI for integration with external beam radiotherapy.

Physics in medicine and biology·2018

A new calibration method uses an electronic portal imaging device (EPID) for precise intensity-modulated radiotherapy (IMRT) dose verification. This technique accurately measures radiation fields, improving treatment accuracy and patient safety.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiotherapy Technology

Background:

  • Accurate dosimetric verification is crucial for intensity-modulated radiotherapy (IMRT).
  • Electronic portal imaging devices (EPIDs) offer potential for in-vivo dosimetry but require robust calibration.
  • Existing methods may lack the spatial resolution and accuracy needed for complex IMRT fields.

Purpose of the Study:

  • To develop and validate a convolution-based calibration procedure for amorphous silicon flat-panel EPIDs.
  • To enable accurate 2-D dosimetric verification of IMRT treatments using EPIDs.
  • To assess the accuracy and spatial resolution of the EPID-based verification method.

Main Methods:

  • Developed a deconvolution algorithm using a two-element scatter kernel (Monte Carlo and empirical) to derive primary fluence from raw EPID images.

Related Experiment Videos

  • Convolved EPID-measured primary fluences with a Monte Carlo dose deposition kernel for phantom dosimetry.
  • Cross-calibrated EPID measurements with ion chamber and film dosimetry for validation.
  • Validated against diamond detector measurements for relative fluence profiles.
  • Main Results:

    • EPID-measured relative fluence profiles showed excellent agreement with diamond detector measurements.
    • EPID-derived dose distributions agreed within 2.1% with film measurements for open fields.
    • Predicted EPID phantom scatter factors (SPE) were within 1% of measured SPE up to 16x16 cm2.
    • Pretreatment IMRT verifications demonstrated good agreement with film, with a mean difference of 0.2 +/- 1.0%.

    Conclusions:

    • The convolution-based calibration procedure enables accurate dosimetric verification of IMRT treatments using EPIDs.
    • The developed method provides high spatial resolution and accuracy suitable for IMRT verification.
    • EPID-based dosimetry offers a reliable and precise alternative for pretreatment IMRT quality assurance.