[Pacemaker syndrome after dual-chamber implantation: consequence of a very long atrial activation delay]
E M Pretorian1, O Jaboureck, P Joly
1Service de cardiologie, centre hospitalier de Douai, route de Cambrai, BP 740, 59507 Douai, France. matei.pretorian@ch-douai.fr
Abstract:
Many circumstances, generally occurring in inappropriate device's adjustment or in specific myocardial conduction's disturbances, can result in dual chamber indications in pace maker syndromes. Our case report is about a man implanted with a dual chamber device with a dilated right atrium and intra atrial conduction delay resulting in a delayed post pacing atrial activation time. The consequence was an atrial contraction occurring during closed atrioventricular valves.