When the blood-back detection system fails: an IABP case report
G H Linley1, E W Bakker, R de Vroege
1Department of Extracorporeal Circulation, Vrije Universiteit Medisch Centrum, Amsterdam, The Netherlands. r.devroege@vumc.nl
Abstract:
A 74-year-old male patient's circulation was supported by an intra-aortic balloon for a period of six days following multiple cardiac infarcts. On the sixth day, several 'leak in IAB circuit' alarms appeared. The nursing staff checked the tubing and refilled the balloon, as described in the manual, without notifying the medical staff or perfusionist. A few hours later, the balloon showed a leak, as blood was seen in the catheter tubing. This leak resulted in considerable damage to the device caused by a large amount of blood migrating back to the tubing and flooding the internal drive system, due to a failing 'blood detection' sensor. The balloon catheter was removed percutaneously and intra-aortic counterpulsation was discontinued. The patient died 20 days later of heart failure. This raises doubts over the adequacy of the protective sensors of such devices and, importantly, how to interpret the present user manuals and the instructions for troubleshooting.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure