Strengths and limitations of commercial tests for hepatitis C virus RNA quantification
Chihiro Morishima1, Minjun Chung, Ka Wing Ng
1Department of Pediatrics, University of Washington School of Medicine, Seattle, Washington, USA.
Journal of Clinical Microbiology
|January 13, 2004
Abstract:
The sensitivity of the Roche COBAS Monitor v.2.0 test was slightly better than that of the Bayer bDNA 3.0 test, but the Monitor test underquantified specimens by 2.5- to 10.6-fold even at relatively low hepatitis C virus RNA concentrations. Dilution prior to assay minimally decreased the reproducibility of the Monitor assay, leading to the recommendation for all specimens to be diluted 1:100 prior to testing by this assay.