Aim:
To develop and evaluate an effective cleaning procedure for rotary nickel-titanium (NiTi) endodontic instruments.
Methodology:
New rotary instruments (ProFile size 25/.04) were contaminated by preparing canals of extracted teeth. Three factors were evaluated to develop an effective cleaning sequence: dry or moist storage before cleaning; mechanical removal (brushing); and chemical dissolution in 1% NaOCl with ultrasonication. Debris on flutes was scored after staining in situ with Van Gieson's solution at x45 magnification. Debris was classified as stained or unstained particulate debris and organic film, and rated as none, slight, moderate or heavy. The effectiveness of a recommended cleaning sequence was tested on different instrument types and in private endodontic practices.
Results:
All new instruments showed metallic spurs and fine particulate debris on the surfaces. After contamination, brushing alone removed most particulate debris, but did not remove organic film. NaOCl effectively removed organic film. Under laboratory conditions, the sequential cleaning procedures (moist storage, brushing followed by immersion in 1% NaOCl and ultrasonic cleaning) totally removed organic debris. Dry storage before cleaning or autoclaving with debris present reduced cleaning effectiveness (P<0.001, one-way ANOVA). In three private practices, the cleaning protocol substantially reduced biological contamination, but complete cleaning was not always achieved (87% clean).
Conclusion:
Complete removal of organic debris from instruments is feasible using a combination of mechanical removal and chemical dissolution, but requires meticulous attention to details.
More Related Videos
Related Concept Videos
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...