Related Experiment Videos
Free haemoglobin interferes with detection of endothelin peptides
1Dept. of Pharmacology, Karolinska Institute, Stockholm, Sweden.
Abstract:
High levels of endothelin-like immunoreactivity were detected in red blood cells from rat, pig and man. When characterized on HPLC the immunoreactivity coeluted with haemoglobin, however. Thus, the high levels of endothelin-like immunoreactivity did not reflect occurrence of endothelin peptides but rather the interference of haemoglobin in the RIA. Free haemoglobin > 0.8 g/l (which may occur in haemolytic samples) increased measured plasma "endothelin-like immunoreactivity". SepPak extraction of plasma samples markedly reduced this interference, although some effect still remained at high haemoglobin concentrations. The influence of microperoxidase in the RIA suggests that the interference is related to the haeme portion of haemoglobin and thus may be extended to other haeme-containing proteins, e.g. cytochrome c or guanylate cyclase. The present findings emphasize the importance of characterizing endothelin-like immunoreactivity with HPLC, especially in haemolytic samples.