Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo
Published on: September 26, 2016
Direct EPR detection of transient and continuous wave signals at 2.5 GHz
Jörg Forrer1, Hansjürg Schmutz, Rene Tschaggelar
1Physical Chemistry, ETH Zurich, CH-8093 Zurich, Switzerland.
Abstract:
Direct detection of free induction decays and electron spin echoes, and the recording of echo-detected EPR spectra and electron spin echo envelope modulation patterns at a microwave frequency of 2.5 GHz is demonstrated. This corresponds to the measurement of the transverse magnetization in the laboratory frame, rather than in the rotating frame as usually done by down-converting the signal (homodyne detection). An oscilloscope with a 6-GHz analog bandwidth, a sampling rate of 20 GigaSamples per second, and a trigger frequency of 5 GHz for the edge trigger and 750 MHz for the advanced trigger, is used in these experiments. For signal averaging a 3-GHz microwave clock divider has been developed to synchronize the oscilloscope with the frequency of the EPR signal. Moreover, direct detection of continuous wave EPR signals at 2.5 GHz is described.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Electronic Distance Measuring Instruments
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...