Related Experiment Video
Updated: Aug 29, 2026

Custom-made Microdialysis Probe Design
Published on: July 21, 2015
Evaluation of an in vitro dialysis system to predict drug removal
Joanna Q Hudson1, Thomas J Comstock, George M Feldman
1Department of Pharmacy, The University of Tennessee, Memphis, TN 38163, USA. jhudson@utmem.edu
Background:
Variation in the extent of drug removal under different dialysis conditions presents a challenge for prediction of drug elimination and dosage regimen adjustment during haemodialysis (HD). Dependence on clinical pharmacokinetic studies in HD patients for dosing guidelines is problematic given the increasing number of dialysers with variable rates of drug removal. Thus, the purpose of this study was to characterize drug removal using an in vitro system and to evaluate its reliability to predict in vivo elimination by HD using vancomycin (VANC) as a model drug.
Methods:
In vitro dialysis was performed for 2 h (volume 4.0 l normal saline, initial VANC concentration 30 mg/l, flow rate 300 ml/min, dialysate flow 800 ml/min) using four different dialysers: polymethylmethacrylate (BK-2.1 U), polysulfone (F-80), AN69 (Filtral-20) and hemophan (COBE 700HE). The in vitro dialysis clearance for VANC (CL(D)) for the polysulfone dialyser was compared with values determined in eight HD patients. In vitro VANC CL(D) for all dialysers was compared with the clearance and KoA for B12 reported for each dialyser.
Results:
In vitro VANC CL(D) values were 93+/-11 ml/min for the polymethylmethacrylate BK-2.1, 136+/-7 ml/min for the AN69, 65+/-9 ml/min for the hemophan COBE 700HE and 143+/-10 ml/min for the polysulfone F80. The CL(D) for the polysulfone F80 was not statistically different from the in vivo CL(D) of 135+/-18 ml/min (P = 0.48). In vitro VANC CL(D) correlated with the B12 CL(D) (r(2) = 0.77) and the B12 KoA (r(2) = 0.63) reported for each dialyser.
Conclusion:
VANC CL(D) in HD patients for the polysulfone dialyser was correctly predicted using the in vitro dialysis system. Use of this system may be superior to estimations of drug CL(D) based on dialyser information provided by the manufacturer for compounds of similar molecular weight.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Drug Elimination by Renal Route: Tubular Secretion