Assessing comfort, safety, and patient satisfaction with three commonly used penile compression devices
K N Moore1, S Schieman, T Ackerman
1University of Alberta Faculty of Nursing and Faculty of Medicine, Edmonton, Alberta, Canada.
Objectives:
To assess the safety, efficacy, comfort, and patient satisfaction with three penile compression devices: the Cunningham clamp, C3, and U-Tex.
Methods:
The devices were tested in random order in a multiple-period, crossover study design using a Latin squares configuration. The subjects had undergone radical prostatectomy 6 months or more before the study, had no neurologic or cognitive impairment, and had not undergone radiotherapy. Baseline penile Doppler ultrasonography was followed by ultrasound scanning with each device. In random order, subjects completed a 4-hour pad test, with and without each device, and the questionnaire.
Results:
Twelve men completed the study. The mean Mini-Mental State Examination score was 29.6 (SD 1.2, range 27 to 30). The mean urine loss at baseline was 122.8 g (SD 130.8). The mean urine loss with each device was 53.3 g (SD 65.7) with the U-Tex, 32.3 g (SD 24.3) with C3, and 17.1 g (SD 21.3) with the Cunningham clamp (P <0.05). No device had an impact on the resistive index; the C3 and U-Tex allowed good cavernosal artery flow, and the Cunningham clamp significantly lowered the distal blood flow velocity (from 12.5 to 7.3 cm/s [left systolic velocity] to 9.5 cm/s [right systolic velocity]) even at the loosest setting. The Cunningham clamp was ranked positively by 10 of 12 men; 2 of 12 men rated the C3 positively; none rated the U-Tex positively.
Conclusions:
The Cunningham device was the most efficacious and most acceptable to users, but also contributed to reduced systolic velocity in all men. None of the devices completely eliminated urine loss when applied at a comfortable pressure. Individualized instruction to cognitively capable men is necessary to ensure appropriate application, comfort, and fit.
Related Concept Videos
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Assessing Blood Pressure