Related Experiment Video
Updated: Jul 11, 2026

08:19
Digital Printing of Titanium Dioxide for Dye Sensitized Solar Cells
Published on: May 4, 2016
Technical note: a procedure for the preparation and quantitative analysis of samples for titanium dioxide
W D Myers1, P A Ludden, V Nayigihugu
1Department of Animal Science, University of Wyoming, Laramie 82071-3684, USA.
Journal of Animal Science
|February 3, 2004
Summary
A new, rapid method accurately analyzes titanium dioxide (TiO2) in feed and fecal samples within 4.5 hours. This procedure offers superior TiO2 recovery compared to traditional dry-ash methods.
Area of Science:
- Analytical Chemistry
- Animal Nutrition
Background:
- Accurate determination of titanium dioxide (TiO2) is crucial for feed digestibility studies.
- Existing dry-ash methods for TiO2 analysis are time-consuming and can yield inaccurate results.
Purpose of the Study:
- To develop a rapid and accurate analytical procedure for titanium dioxide (TiO2) quantification.
- To compare the efficacy of the new method against a conventional dry-ash procedure.
Main Methods:
- Samples were digested using concentrated sulfuric acid (H2SO4) and hydrogen peroxide (H2O2).
- Absorbance was measured at 410 nm to determine TiO2 concentrations.
- A linear standard curve was established using spiked blanks.
Main Results:
- The developed procedure provided complete analysis within 4.5 hours.
- TiO2 recoveries averaged 96.7-98.5% across different organic matter sources.
- The new method significantly outperformed the dry-ash procedure, which yielded recoveries of 53.1-83.8%.
Conclusions:
- The novel digestion and spectrophotometric method is a rapid and accurate alternative for TiO2 determination.
- This procedure enhances the reliability of using TiO2 as an indigestible marker in animal nutrition studies.
More Related Videos
Related Concept Videos
Sample Preparation for Analysis: Overview
Sample preparation is an essential step in the analytical process. It involves preparing a sample so that it can be analyzed accurately. The goal is to extract the analyte, the substance you want to measure, from the sample while removing any components that may interfere with the analysis. Sample preparation techniques vary depending on the physical state of the sample.
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Sample Preparation for Analysis: Advanced Techniques
Accurate analysis of complex samples often requires advanced preparation techniques to achieve reliable and reproducible results. Samples containing inorganic or organic materials can be challenging to dissolve or decompose effectively. Standard sample preparation methods include acid digestion, fusion, dry ashing, and wet digestion.
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...