Related Experiment Video
Updated: Aug 29, 2026

Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
Published on: December 20, 2016
Coupled ion chromatography for the determination of chloride, phosphate and sulphate in concentrated nitric acid
Magdalena Biesaga1, Nicole Schmidt, Andreas Seubert
1Laboratory for Flow Analysis and Chromatography, Department of Chemistry, Warsaw University, Pasteura 1, Warsaw 02-093, Poland.
Abstract:
A coupled ion chromatography (IC) system was used for the determination of chloride, sulphate and phosphate in high-purity nitric acid. Such a high ionic strength matrix causes a selectivity problem in single IC systems. The first part of the system is used for a pre-separation of the analytes from the nitrate matrix. A specially designed high-capacity anion exchanger with low retention for the analytes and high retention for nitrate was developed. The eluent stream containing the analytes was transferred to the second part of the system via a heart-cut valve and a pre-concentration column. The second system utilizes a high performance anion exchanger and is used to quantify the analytes. Recoveries of the analytes are 80-100% for phosphate, and around 100% for sulphate and chloride. Detection limits for chloride, sulphate and phosphate in concentrated nitric acid (69% w/w) are 0.1, 1 and 5 mg/l, respectively.
More Related Videos
Related Concept Videos
Ion-Exchange Chromatography
Precipitation Titration Curve: Analysis
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Qualitative Analysis
For instance, group IV...
Precipitation Titration: Overview
A precipitation titration curve demonstrates the change in concentration of the titrant or analyte upon adding the...