Minimizing false positive diagnoses in newborn screening for galactosemia
D M Frazier1, E H Clemons, H N Kirkman
1Brain and Development Research Center, Chapel Hill, North Carolina.
Summary
Heat and humidity degrade galactose-1-phosphate uridyltransferase activity in mailed blood spots, leading to false positive galactosemia diagnoses. Proper drying and sealing of blood spots are crucial for accurate enzyme activity testing.
Area of Science:
- Biochemistry
- Clinical Diagnostics
- Newborn Screening
Background:
- Galactosemia diagnosis relies on measuring galactose-1-phosphate uridyltransferase (GALT) activity in dried blood spots.
- Environmental factors like heat and humidity can affect GALT activity, potentially leading to inaccurate results.
- False positive diagnoses of galactosemia can cause unnecessary parental anxiety and further testing.
Purpose of the Study:
- To investigate the impact of heat and humidity on GALT activity in mailed blood spots.
- To identify optimal methods for preserving GALT activity during sample transport and storage.
- To differentiate between true galactosemia and sample deterioration.
Main Methods:
- Blood spots were exposed to varying conditions of heat and humidity.
- Galactose-1-phosphate uridyltransferase activity was measured fluorometrically.
- Phosphoglucomutase activity was simultaneously determined to assess sample integrity.
- Different drying and sealing methods for blood spots were evaluated.
Main Results:
- Both heat and humidity significantly reduce GALT activity in blood spots.
- Allowing blood spots to dry completely before sealing in plastic bags preserves GALT activity.
- Sealing moist blood spots in plastic bags exacerbates GALT activity loss.
- Fluorometric assays with simultaneous phosphoglucomutase determination accurately distinguish between sample deterioration and true galactosemia.
Conclusions:
- Heat and humidity are primary causes of reduced GALT activity in mailed blood spots, leading to false positive galactosemia diagnoses.
- Proper sample handling, including complete drying and appropriate sealing, is essential for reliable GALT activity testing.
- Simultaneous measurement of phosphoglucomutase activity provides a critical control for sample quality in galactosemia screening.