Dosimetry of shaped electron fields using a radial integration method.

R Muller-Runkel1

  • 1Saint Margaret Mercy Health Care Centers, Oncology Center, Hammond, IN 46320.

Medical Dosimetry : Official Journal of the American Association of Medical Dosimetrists
|January 1, 1992
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Measurement of dosimetric parameters for the Alpha-Omega high-dose-rate Iridium-192 source.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2005
Same author

Changing face and different countenances of prostate cancer: racial and geographic differences in prostate-specific antigen (PSA), stage, and grade trends in the PSA era.

International journal of cancer·2001
Same author

Determination of the tissue attenuation factor along two major axes of a high dose rate (HDR) 192Ir source.

Medical physics·1999
Same author

Validity of the interval method for the determination of the anisotropy factor of high dose rate 192Ir sources.

International journal of radiation oncology, biology, physics·1997
Same author

Evaluation of a commercial three-dimensional electron pencil beam algorithm.

Medical physics·1997
Same author

Anisotropy measurements of a high dose rate Ir-192 source in air and in polystyrene.

Medical physics·1994
Same journal

Predictive dosimetric parameters for adaptive replanning in proton beam therapy for thoracic cancer.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

A case study of inhale and exhale split-course VMAT treatment to minimize dose to ribs.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Prospective clinical implementation of the robust optimization planning technique in long-course photon radiotherapy treatment for lung cancer.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Dosimetric comparison of target coverage, dose gradient, and low-dose bath in left-sided accelerated partial breast irradiation: CyberKnife versus conventional linac-based noncoplanar VMAT and 3D-CRT.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Cross-prescription validation of knowledge-based planning in prostate VMAT using a structured validation framework.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Dosimetric evaluation of beam block techniques in helical tomotherapy for extranodal nasal-type NK/T-cell lymphoma.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
See all related articles

An analytical method for calculating monitor units in shaped electron fields shows high accuracy. This approach is feasible for clinical applications, with most treatments showing excellent dose agreement.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Dosimetry

Background:

  • Accurate calculation of monitor units (MUs) is crucial for precise electron beam radiotherapy.
  • Shaped electron fields present unique challenges for traditional dosimetry calculations.

Purpose of the Study:

  • To investigate the feasibility of an analytical approach for calculating MUs in shaped electron fields.
  • To evaluate the accuracy of this method in clinical scenarios.

Main Methods:

  • A radial integration method was employed to compute dose at prescription depth.
  • The analytical concept was implemented in a commercial treatment planning system.
  • Validation was performed using arbitrary fields and 66 clinically relevant shaped electron fields.

Related Experiment Videos

Main Results:

  • 71% of clinical fields showed measured and prescribed dose agreement within 3.5%.
  • 91% of fields achieved agreement within 5.5%.
  • The greatest discrepancy of -7.2% was observed for a specific internal mammary field.

Conclusions:

  • The analytical approach for calculating MUs in shaped electron fields is feasible and accurate.
  • This method demonstrates good clinical applicability for various treatment sites.
  • Further refinement may be needed for specific complex field geometries.