Related Experiment Videos
A comparison of round-window and transtympanic promontory electric stimulation in cochlear implant candidates
P R Kileny1, T A Zwolan, S Zimmerman-Phillips
1Department of Otolaryngology, University of Michigan Medical Center, Ann Arbor.
Abstract:
We compared within-subjects electrical thresholds and dynamic ranges obtained with direct round-window and transtympanic promontory stimulation carried out preoperatively in 12 patients who were candidates for a cochlear implant. Square waves with frequencies of 50, 100, 200, and 400 Hz were delivered in a 50% duty cycle to both sites in each patient. With the exception of threshold at 50 Hz (promontory thresholds were lower than round-window thresholds), there were no statistically significant differences for either thresholds or dynamic ranges between the two sites of stimulation. There was a general trend for round-window thresholds to be lower and dynamic ranges larger, especially for the higher frequencies of stimulation. Mean threshold slopes for the two sites of stimulation were nearly identical.