Measuring Pressure Volume Loops in the Mouse
Published on: May 2, 2016
Dynamic elastic pressure-volume loops in healthy pigs recorded with inspiratory and expiratory sinusoidal flow
Ulrika Bitzén1, Björn Drefeldt, Lisbet Niklason
1Department of Clinical Physiology, Lund University Hospital, 221-85 Lund, Sweden. Ulrika.Bitzen@klinfys.lu.se
Objective:
The objective was to analyse relationships between inspiratory and expiratory static and dynamic elastic pressure-volume (P(el)/V) curves in healthy pigs.
Design:
The modulated low flow method was developed to allow studies also of the expiratory P(el)/V curves. Static P(el)/V (P(el,st)/V) and dynamic P(el)/V (P(el,dyn)/V) loops were studied in healthy pigs.
Setting:
Animal research laboratory in a university hospital.
Material:
Ten healthy anaesthetised and paralysed pigs.
Interventions And Measurements:
A computer controlled a Servo Ventilator 900C with respect to respiratory rate, inspiratory flow and expiratory pressure to achieve a sinusoidal modulation of inspiration and expiration for determination of P(el,dyn)/V loops from zero end-expiratory pressure (ZEEP) and from a positive end-expiratory pressure (PEEP) of 6 cmH(2)O to 20, 35 and 50 cmH(2)O. The same system was used for studies of P(el,st)/V loops with the flow-interruption method from ZEEP and PEEP to 35 cmH(2)O. Recordings were analysed with an iterative technique.
Results:
The feasibility of automated determination of P(el,dyn)/V loops was demonstrated. Differences between P(el,dyn)/V and P(el,st)/V loops were explained by viscoelastic behaviour. P(el,st)/V loops recorded from PEEP to 35 cmH(2)O showed no significant hysteresis, indicating a non-significant surface tension hysteresis. P(el,dyn)/V loops from PEEP and both P(el,st)/V and P(el,dyn)/V loops from ZEEP to 35 cmH(2)O showed hysteresis. This indicates that lung collapse/re-expansion caused P(el)/V loop hysteresis which, in P(el,dyn)/V loops, was augmented by viscoelastic behaviour.
Conclusions:
Viscoelasticity influences P(el,dyn)/V curves. Hysteresis caused by surface tension merits re-evaluation. Lung collapse and re-expansion may be indicated by hysteresis of P(el)/V loops.