Impact of strut thickness on late luminal loss after coronary artery stent placement

Saskia Z H Rittersma1, Robbert J de Winter, Karel T Koch

  • 1Department of Cardiology, Academic Medical Center, Amsterdam, The Netherlands.

Insights

Thinner coronary artery stent struts (50 microm) significantly reduced late luminal loss compared to thicker struts (≥90 microm). Strut thickness is an independent predictor of late luminal loss after stenting.

Area of Science:

  • Cardiovascular Medicine
  • Interventional Cardiology
  • Biomedical Engineering

Background:

  • Coronary artery disease necessitates stenting to restore blood flow.
  • Stent design, particularly strut thickness, may influence long-term outcomes.
  • Late luminal loss is a key metric for assessing in-stent restenosis and stent performance.

Purpose of the Study:

  • To investigate the impact of coronary artery stent strut thickness on angiographic late luminal loss.
  • To determine if stent strut thickness is an independent predictor of late luminal loss.

Main Methods:

  • Single-center observational cohort study.
  • Inclusion of 663 patients receiving ACS Multilink stents in native coronary vessels.
  • Angiographic follow-up at 6-10 months to measure late luminal loss.

Main Results:

  • Patients treated with thin-strut stents (50 microm) showed significantly less late luminal loss (0.92 ± 0.59 mm) than those with thick-strut stents (≥90 microm) (1.06 ± 0.71 mm).
  • Multivariate regression analysis identified strut thickness as an independent predictor of late luminal loss (p=0.011).

Conclusions:

  • Coronary artery stent strut thickness is a significant factor influencing late luminal loss.
  • Thinner strut stents are associated with better angiographic outcomes, reducing late luminal loss.
  • These findings support the consideration of strut thickness in stent selection for coronary interventions.