Scatter correction on its own increases image contrast in Tl-201 myocardium perfusion scintigraphy, but does it also
Koos J A K Blokland1, Wouter H de Vos tot Nederveen Cappel, Berthe L F van Eck-Smit
1Leiden University Medical Centre, Department of Radiology, Division of Nuclear Medicine, Leiden, The Netherlands. j.a.k.blokland@lumc.nl
Unlabelled:
Poor and variable spatial resolution of the gamma camera, the movement of the heart and, above all, the inclusion of scattered photons in the acquisition data contribute to the deterioration of image contrast in 201Tl myocardium perfusion studies. Scatter correction algorithms may correct for the latter factor by removing (most of) the scattered photons from the acquisition data.
Methods:
In this study we investigated the contrast changes induced by the Triple Energy Window scatter correction method (TEW) applied to clinical 201Tl myocardium perfusion studies and its influence on the reading of the images. Stress and rest studies of 30 consecutive patients were used for this study. Maximum image contrasts were measured between the myocardium and the left ventricular cavity in four mid-ventricular short axis slices, as well as between normally and abnormally perfused myocardium using bull's-eye displays of the activity within the myocardium. To assess image quality and perfusion abnormalities, an experienced nuclear medicine physician, blind to patient characteristics, visually reviewed all studies.
Results:
In all individual measurements, the maximum contrast after scatter correction was higher than without correction (p < 0.001). The average increase in contrast between the myocardium and the left ventricular cavity was 43% and 48% for stress and rest studies respectively. The contrast within the myocardium increased by 25% and 32% respectively. After TEW, image quality was rated lower in almost half of the studies, while in only one study the quality was rated higher. In stress studies 11 additional perfusion defects were observed, with rest studies revealing 15 more defects after TEW, but this difference was not significant. Cohen's kappa indicated a moderate agreement of the image reading between studies with and without scatter correction.
Conclusion:
We conclude that image contrast improves significantly by scatter correction. However, image quality decreased as a result of an unfavorable signal-to-noise ratio. As an overall result, no significant change in the clinical outcome of the studies could be shown. Additional training of the readers may be required to obtain optimal results.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...