Published on: December 18, 2016
Optimisation of T2 and M0 measurements of bi-exponential systems
Anastasios Anastasiou1, L D Hall
1Herchel Smith Laboratory for Medical Chemistry, University of Cambridge, School of Clinical Medicine, Robinson Way, Cambridge, England, CB2 2PZ, United Kingdom. T.Anastasiou@cantab.net
Abstract:
Cramer-Rao theory and computer simulations were used to show that the errors involved in calculating the magnetization and relaxation parameters of a two-component system decrease with: (1) increasing SNR, (2) increasing number of echoes used in the fitting procedure, and (3) increasing ratio of the relaxation times of the two components, T(22)/T(21). Images of bi-compartmental phantoms of known T(2) values were acquired using an optimized imaging sequence, and an optimized fitting algorithm was used to calculate the T(2) values of the two components by fitting the resulting images to a bi-exponential decay model. Accuracy better than 6% was achieved in the calculations of the T(2) values of the two components, and region fitting provided better accuracy than pixel-by-pixel fitting. The procedures were used to calculate the T(2) and M(0) values of equine carpal bones with known degree of radiographic bone sclerosis. Although the T(2) and M(0) values of both water and fat components all decreased with the degree of radiographic bone sclerosis, the transverse relaxation of the water component, T(2W), showed a greater decrease with advanced stages of bone sclerosis.
Related Concept Videos
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
Atomic Nuclei: Nuclear Spin State Population Distribution
NMR Spectrometers: Resolution and Error Correction
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations