Multicenter trial for the set-up of a MRI quality assurance programme
P Colombo1, A Baldassarri, M Del Corona
1Health Physics Department, Ospedale Niguarda Cà Granda, Milano, Italy. sfsnig@mailserver.unimi.it
Abstract:
Many international protocols related to RMI-QC program are focused on acquisition methods and analysis of several image quality parameters but rarely normality ranges or measurement frequencies are presented. To address this problem we investigated the variability of many magnetic resonance imaging (MRI) systems with the set-up of multicenter trial. The trial was set up to investigate short-and mid-term variability of two fundamental nongeometric image quality parameters: signal-to-noise (SNR) and integral percent uniformity (U%). Ten centers (12 devices) participated to data collection consisting of a three-step-protocol. First, 10 consecutive images of a phantom were collected with a spin echo sequence. As second step the series collection was repeated 24 h later. Finally a single image acquisition was performed twice a week for 5 weeks. The analysis of results allowed us to define a "physiological" variability of +/-3% of the reference level for both parameters and to conclude that a weekly measurement is adequate to detect relevant variations of device performance.
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI