Background:
Transvascular catheter-based intracardiac echocardiography has been successfully used to help guide catheter ablation and electrophysiologic procedures. It has recently been demonstrated that catheters can be safely placed into the pericardial space to allow for epicardial cardiac mapping and ablation. We evaluated the feasibility of catheter-based intrapericardial echocardiography (IPE) during such procedures to identify cardiac structures and visualize intracardiac catheters.
Methods:
IPE was performed in 7 goats by placing a phased-array ultrasound transducer contained within a 10F steerable catheter into the pericardial space using the same transthoracic subxyphoid approach as used to map and ablate epicardial ventricular tachycardia. Images were obtained of cardiac structures and of intracardiac ablation catheters. After the procedure, the hearts were harvested to assess for possible IPE-related lesions.
Results:
The IPE catheter could be easily placed inside the pericardial space in all animals. In 7 of 7 cases, longitudinal and short-axis views of right- and left-sided chambers and valves were obtained, similar in orientation to transesophageal echocardiography. Visualization of atrial appendages (6/7), pulmonary veins (6/7), coronary arteries (6/7), and coronary sinus (3/6) was also feasible. Assessment of intracardiac transvalvar and venous blood flow was achieved by spectral and color Doppler. The ablation catheter could be clearly visualized inside cardiac chambers. No arrhythmias were induced with IPE catheter manipulation. After harvesting the hearts, no lesions resulting from the procedure were observed.
Conclusion:
In this experimental setting, IPE was able to provide detailed images of cardiac structures and establish the relative position of the ablation catheter.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization III: Left Heart Catheterization
Acute Coronary Syndrome III: Diagnostic Studies