Related Experiment Video
Updated: Aug 26, 2026

An Open-Source Normothermic Perfusion System Designed for Research Scientists
Published on: July 18, 2025
Super high flux hemodialysis at high dialysate flows: an ex vivo assessment
1Department of Intensive Care, Austin & Repatriation Medical Centre, Melbourne, Australia.
Background And Objectives:
The removal of cytokines by standard hemofiltration is limited. Super high flux membranes may significantly improve removal even when used in dialysis mode. We sought to measure cytokine clearance using a large surface super high-flux membrane and a standard hemodialysis setting.
Setting:
ICU laboratory of a tertiary institution.
Subjects:
Six healthy volunteers.
Methods:
Blood form healthy volunteers was incubated for 4 hours with E. coli endotoxin to stimulate cytokine production. Cytokine containing blood was then circulated through a dialysis circuit at 3 different dialysate flow rates. Blood and dialysate were sampled for cytokine and albumin measurements and calculation of clearances.
Results:
Super high-flux dialysis achieved high median cytokine clearances (IL-1 clearance of 106 ml/min, IL-6 clearance of 66.8 ml/min, IL-8 clearance of 61.7 ml/min and TNF clearance of 36.1 ml/min). Increasing dialysate flow rate from 300 to 500 ml/min did not significantly increase cytokine clearances. Albumin clearances however were between 2.7 and 5.4 ml/min.
Conclusions:
Cytokine dialysis is feasible at high dialysate flow rates yielding high cytokine clearances. Albumin loss, however, is appreciable and may require separate supplementation in the clinical setting.