Related Experiment Video
Updated: Aug 26, 2026

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Primary and secondary measurements of 222Rn
1Physikalisch-Technische Bundesanstalt, Bundesallee 100, Braunschweig 38116, Germany.
Abstract:
In the past few years, a primary measurement system has been established at the PTB allowing absolute measurements of 222Rn. The activity can be determined by counting the emitted alpha particles in a defined solid angle after condensing the radon at a cold point under vacuum conditions. This primary measurement system is connected to a solid 226Ra source, where 222Rn can be produced up to an activity of about 2MBq. After the measurement, radon can be completely transferred into a glass bulb or a stainless steel cylinder for use as a transfer standard. The results of the measurements, the uncertainties and the realization of the primary measurement system are described. For quality assurance and fast determination of radon activity, a secondary measurement system with an NaI(Tl) detector is also used. Together, both systems provide a reliable tool for the production of gaseous 222Rn activity standards with relative uncertainties of less than 1%.