Related Experiment Video
Updated: Aug 17, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Calculation of 18F, 99mTc, 111In and 123I calibration factor using the penelope ionization chamber simulation method
1BNM-CEA/LNHB, Laboratoire National Henri Becquerel, Gif sur Yvette Cedex (91191), France. marie-noelle.amiot@cea.fr
Abstract:
A new method using the Monte Carlo code PENELOPE for ionization chamber simulation has already been successfully used for calculating calibration factors needed for the measurements of radionuclides with photon emission (2003, Appl. Radiat. Isot., to be published). This work has been continued at (Laboratoire National Henri Becquerel) in order to calculate the calibration factors for radionuclides with short half-lives used in medical services. Activity measurements of 18F, 99mTc,111In and 123I using the calculated calibration factors were obtained with standard uncertainties equal to 0.6% for 18F, 99mTc and 1.5% for 111In and 123I.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021