Related Experiment Video
Updated: Aug 18, 2026

A Morphometric and Cellular Analysis Method for the Murine Mandibular Condyle
Published on: January 11, 2018
[A study of the magnification of mandible pantomography image]
1Department of Oral Radiology, Suzhou Stomatological Hospital, Jiangsu 215005, China.
Objective:
To study the degree of magnification of mandible pantomography image.
Methods:
The pantomography of 50 mandible specimens marked with lead pellets were taken by TROPHY XL-90C. The distances between each markers on film and in specimen were measured.
Results:
The mean magnification(MM) of mandible pantomography image with TROPHY XL-90C was 1.22. MM of horizontal axle was 1.14. MM of vertical axle was 1.29. MM of horizontal axle was greatest (1.24) in the body of mandible,and lest (1.08) in genion. MM of vertical axle is similar among each regions. MM of vertical axle was greater than that of horizontal axle in all regions.
Conclusion:
The MM parameters of mandible pantomography image were useful to estimate the size of mandible lesions and provided a basis for dental implant more exactly.