The use of a bacteria detection system to evaluate bacterial contamination in PLT concentrates
1Department of Pathology and Laboratory Medicine, Division of Hematology and Transfusion Medicine, The Ottawa Hospital, 1053 Carling Avenue, Ottawa, Ontario K1Y 4E9, Canada. grock@ottawahospital.on.ca
Background:
Random-donor PLTs (RDPs) are functional at 7 days. Nevertheless, since the mid-1980s, concern for bacterial contamination has caused the storage period to be reduced to 5 days. The ability of a bacteria detection system (BDS, Pall) to determine bacterial contamination and permit extension of the PLT shelf life to 7 days was assessed.
Study Design And Methods:
Blood was collected into CP2D and leukoreduced RDPs were prepared. Upon arrival at the hospital, a 2- to 3-mL aliquot was removed from each RDP and introduced into the Pall BDS pouch with a sterile docking device. The pouch was incubated at 37 degrees C for 24 hours and then the oxygen content was measured to determine bacterial contamination. Additionally, the RDPs were pooled and an aliquot was removed for culture with standard manual techniques. CCIs were calculated 1 hour after infusion.
Results:
A total of 12,062 individual RDPs were tested. The Pall BDS detected bacteria in 5 units. All of these were positive on repeat sampling. Propionibacterium acnes, coagulase-negative Staphylococcus, and Bacillus species were confirmed by manual technique in 3 units, one could not be identified, and one was negative. Aliquots from PLT pools were positive in 80 of 2201 pools when tested by manual methods. Of these, 79 were false-positives and 1 unit contained coagulase-negative Staphylococcus. The Pall BDS was easy to use and required less than 5 minutes for all manipulations. After 7 days of storage, the PLTs gave an average CCI of 16 x 10(11)+/- 3.39 x 10(11) 1 hour after transfusion (n = 9).
Conclusions:
The Pall BDS permits evaluation of RDPs for bacterial contamination. Culture-negative PLTs were successfully transfused in our institution up to and including 7 days after storage with good CCIs.