Related Experiment Video
Updated: Aug 26, 2026

Using a Knee Arthrometer to Evaluate Tissue-specific Contributions to Knee Flexion Contracture in the Rat
Published on: November 9, 2018
Brief report: validation of a system for automated measurement of knee laxity
M T Thompson1, M A Conditt, S K Ismaily
1Institute of Orthopedic Research and Education, 6550 Fannin Street, Suite 2512, Houston, TX 77030, USA. mthompso@bcm.tmc.edu
Objective:
To determine the accuracy and repeatability of an automated quantitative fluoroscopic imaging system for measuring knee laxity.
Design:
Cadaveric validation study.
Background:
Current methods of measuring anterior-posterior laxity lack sufficient accuracy and repeatability. A commercially developed fluoroscopic software package, capable of measuring laxity, required validation.
Methods:
Five human cadaveric knees were used. A constant force of 130 N was applied anteriorly and posteriorly in turn to the tibia of each knee with the femur fixed in 30 degrees and 90 degrees of flexion. Quantitative fluoroscopic measurements of anterior-posterior laxity were determined using image analysis software. Fluoroscopic results were compared to the true anterior-posterior displacements of the tibia, which were simultaneously recorded using linear transducers directly attached to the cadaveric specimens.
Results:
The quantitative fluoroscopic method underestimated laxity by an average of 0.40 mm with a root mean square error of 0.49 mm. The 95% confidence intervals for anterior and posterior laxity error were calculated to be -0.99 to 0.25 mm and -0.89 to 0.03 mm, respectively, where a negative error represents an underestimation.
Conclusions:
The quantitative fluoroscopic method offers a dramatic improvement in accuracy over current laxity measurement techniques and acceptable repeatability for assessing ligament damage.
Relevance:
The considerably more accurate, validated measurement system of this study could improve ligament assessment and diagnosis, and the recognition of injuries otherwise undetected with current methods.