Related Experiment Video
Updated: Aug 25, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Evaluation of the oesophageal-tracheal double-lumen tube (Combitube) during general anaesthesia
Werner Rabitsch1, Peter Krafft, Franz X Lackner
1Department of Internal Medicine I, Intensive Care Unit, University of Vienna, Vienna, Austria.
Objectives:
Evaluation of safety and effectiveness of the Combitube during general anaesthesia.
Patients And Methods:
250 patients undergoing general anaesthesia were enrolled in the study. The respective types and duration of surgery, ease of insertion of the Combitube, and potential complications were recorded. Maximum ventilatory pressures and leak fraction were also evaluated in this study.
Results:
Duration of surgery varied between 20 and 410 min. More than 96% of the blind Combitube insertions were successful at the first attempt, with a mean time of less than 18 +/- 5 seconds (range 12-24 seconds). In 99% of patients the Combitube worked well, and adequate oxygenation and ventilation was possible. All patients were haemodynamically stable during the entire duration of surgery. In all patients, pulse oximetry showed an oxygen saturation of 97 +/- 2% and an end-tidal carbon dioxide of 38 +/- 6 mmHg. Leak fraction, calculated as a fraction of the inspired volume, did not increase to more than 5% up to a ventilation pressure of 40 cm H2O. Superficial laceration occurred in 18 patients (7.2%) without further sequelae. No severe injuries were observed during the study period.
Conclusion:
Ventilation via the Combitube appears to be safe and effective during general anaesthesia. Practice in elective cases is a requirement for successful use in an emergency situation.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy