Validation of two algorithms to evaluate the interface between bone and orthopaedic implants
Related Experiment Video
Updated: Aug 25, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Debora Testi1, Monica Simeoni, Cinzia Zannoni
1Laboratorio di Tecnologia Medica, Istituti Ortopedici Rizzoli, Via di Barbiano, 1/10, 40136 Bologna, Italy. testi@tecno.ior.it
Abstract:
The level of fit and fill of the prosthetic stem in the femoral canal is an important parameter when planning a cementless total hip arthroplasty. However, the standard templates used in combination with radiographs are not always effective in the pre-operative evaluation of the level of fitting. For this reason, two algorithms were developed able to provide clinically relevant three-dimensional indicators of the implant fit and fill in the host femur, based on the CT data of each specific patient as collected in vivo. In this study the computational methods were described and validated using digital phantom datasets. Then the algorithms were applied for in vivo datasets and the sensitivity of each indicator was evaluated. The validation showed that the two algorithms are accurate from a computational point of view. Moreover, the in vivo testing demonstrated that the developed methods provide reasonable quantitative indicators of the stem positioning in the femoral CT dataset.