Clinical evaluation of an automated oscillometric blood pressure wrist device
Alexandra C Cuckson1, Paul Moran, Paul Seed
1Maternal and Foetal Research Unit, GKT School of Medicine, St Thomas' Hospital, London, UK.
Objectives:
Up to one-third of all blood pressure devices sold are for measurement at the wrist. Wrist monitors are thought to be subject to errors associated with movement and incorrect posture. We used the British Hypertension Society protocol to determine the accuracy of the Omron Rx automated wrist blood pressure monitor in adults. We also investigated the errors associated with miscuffing and posture with measurements at the wrist compared to the upper arm.
Methods:
The study was conducted at a large teaching hospital in the UK. Participants 18 years and over were eligible for inclusion. Exclusion criteria were cardiac arrhythmias and inaudible Korotkoff sounds. Eighty-five participants were selected for the main validation study. Thirty-four subjects took part in the study investigating miscuffing issues and 15 subjects were included in the study investigating postural errors.
Results:
The Omron Rx achieved an overall grade B according to the British Hypertension Society Protocol. The mean differences (standard deviation) between the observers and the device were 0.3 (9.0) and 2.5 (8.8) mmHg for systolic and diastolic blood pressure respectively. In the medium and high-pressure ranges (> or =130/80 mmHg) the device achieved a grade C. When used at home the mean blood pressure reading over 60 measurements in 15 subjects was 123/75 mmHg when taken with the Omron Rx and 124/77 mmHg when taken with the automated upper arm device with no significant difference in the standard deviation of the readings. When compared with mercury sphygmomanometry using both a medium and large cuff, the wrist device readings were closest to the mercury measurements when the most appropriately sized cuff was used.
Conclusions:
The Omron Rx wrist monitor is accurate according to the British Hypertension Society protocol, but narrowly fails the Association of the Advancement of Medical Instrumentation criteria. Measurements at the wrist are not associated with miscuffing errors. Use of the wrist device by patients in the home setting gives comparable readings to a standard upper arm device.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure