Abnormal leukocyte histogram with a factitious automated white cell count due to Plasmodium vivax parasitemia--a case
Naveen Kakkar1, Sumeet Guliani
1Department of Pathology, Christian Medical College & Hospital, Ludhiana, Punjab. n_kakkar@satyam.net.in
Abstract:
Automated hematology analyzers are designed to produce accurate and precise blood counts on specimens which are either normal or show only numerical abnormalities. However they alert the operator in samples with unusual characteristics which could result in inaccurate measurement or require review of the blood smear. Even the most sophisticated instruments cannot recognize all the significant abnormalities that can be recognized by a human observer. Spurious automated white cell counts could result from platelet aggregates, nucleated red cells, non lysis of cells and a variety of other causes. We present a case with spurious leukocytosis due to Plasmodium vivax parasitemia which was flagged by the automated analyzer.
Related Concept Videos
Disorders of Leukocytes
Leukopenia may result from bone marrow disorders, autoimmune diseases, and infectious diseases. For example, conditions such as multiple myeloma and aplastic anemia can impair the bone marrow's ability to produce adequate leukocytes. Similarly, autoimmune diseases like lupus and viral infections such as HIV can prompt the immune system...