Related Experiment Video
Updated: Aug 25, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Reliability of the IOLMaster in measuring corneal power changes after photorefractive keratectomy
Nicola Rosa1, Luigi Capasso, Michele Lanza
1Department of Ophthalmology, 2nd University of Naples, Naples, Italy. nicola.rosa@unina2.it
Purpose:
To test the accuracy of the IOLMaster (Carl Zeiss) in detecting corneal power changes after photorefractive keratectomy (PRK).
Setting:
Department of Ophthalmology, 2nd University of Naples, Naples, Italy.
Methods:
Two hundred twenty-five consecutive eyes that had PRK (mean -5.13 diopters [D] +/- 2.98 [SD] [range +0.25 to -16.25 D]) were analyzed. The data included preoperative and postoperative (1, 3, and 6 months) subjective refraction and computerized keratometry. Statistical analysis was performed to determine the correlation between the changes in the subjective refraction at the corneal plane and the changes in keratometry.
Results:
The mean difference between the changes in refraction and the measured corneal changes was 0.75 +/- 1.13 D (range -3.84 to +7.68 D) at 1 month, 0.92 +/- 1.10 D (range -0.87 to +7.93 D) at 3 months, and 0.75 +/- 0.98 D (range -1.70 to +3.85 D) at 6 months. The difference was significant (P<.001).
Conclusion:
Automated keratometry provided by the IOLMaster did not accurately reflect the effective refractive changes after PRK, particularly in eyes that had a high dioptric treatment.