Related Experiment Video
Updated: Aug 25, 2026

A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Development and validation of a high-performance liquid chromatographic method for the determination of buspirone in
1Division of Pharmaceutical Chemistry, School of Pharmacy, University of Athens, Panepistimiopolis, Zografou 157 71, Athens, Greece.
Abstract:
A stability indicating, reversed-phase high-performance liquid chromatographic method was developed and validated for the determination of buspirone (Bsp) in pharmaceutical dosage forms. The use of a semi-micro XTerra MS C18 (150 mm x 3.0 mm i.d., 5 microm particle size) analytical column, results in substantial reduction in solvent consumption and increased sensitivity. The mobile phase consisted of a mixture of 0.010 M ammonium acetate (pH 4.0) and methanol (55:45, v/v), pumped at a flow rate 0.30 ml min-1. The UV detector was operated at 245 nm. The retention times for lidocaine (Ldc), which was used as internal standard, and buspirone were 4.57 and 7.72 min, respectively. The calibration graph was ranged from 1.00 to 5.00 microg ml-1, while detection and quantitation limits were found to be 0.22 and 0.67 microg ml-1, respectively. The intra- and inter-day relative standard deviation (% R.S.D.) values were less than 1.94%, while the relative percentage error (% Er) was less than 4.0% (n = 5). The method was applied to the quality control of commercial tablets and content uniformity test and proved to be suitable for rapid and reliable quality control.
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors