Related Experiment Video
Updated: Aug 6, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading
G Cheung1, P Zalzal, M Bhandari
1Division of Orthopaedic Surgery, Mount Sinai Hospital, 600 University Avenue, Suite 476A, Toronto, Ont., Canada M5G 1X5.
Abstract:
Intramedullary nails are routinely used in the treatment of fractures of the femur. While their effectiveness has been demonstrated clinically, a number of complications, including bone refracture and implant failure, persist. This paper presents novel three-dimensional finite element (FE) models, at four stages of gait, of: (i) a realistic femur analogue known as third generation composite bone, and (ii) a system consisting of an intramedullary nail implanted in the femur of (i). A comparison of experimentally measured strains on the surface of the femur with those predicted by the FE model revealed good agreement. The models were then used to identify implant/bone load sharing patterns, and areas of stress concentration in both the intramedullary nail and the bone, when statically locked by one or two screws at either end. The results of this study can be used to guide future implant design and surgical procedure.