[Cuff pressure monitoring and regulation in adults]
O Kunitz1, R Jansen, E Ohnsorge
1Klinik für Anästhesiologie, Universitätsklinikum der RWTH Aachen. oliver.kunitz@mutterhaus.de
Objectives:
In adults the use of cuffed endotracheal tubes is the standard technique. Nitrous oxide increases the cuff pressure secondary to diffusion through the cuff membrane. The aim of the study was to verify a newly designed cuff pressure regulating device and comparison of postoperative complications.
Methods:
In a prospective, randomized, open trial the cuff pressure and the incidence of postoperative complications (e.g., hoarseness, coughing and pain while swallowing) were measured using the newly designed automatic pressure monitoring and regulating device Cuff Pressure Control (Tracoe) and a conventional handheld manometer. A total of 40 patients were assigned to the automatic group with a goal of 25.5 cmH(2)O and 40 patients to the conventional group where the pressures were kept between 20 cmH(2)0 and 40 cmH(2)0 with intermittent measurements and manual pressure release.
Results:
The automatic device reliably maintained the pressure at the chosen constant level within +/-2 cmH(2)O. In the control group increases in cuff pressure to 40 cmH(2)O were common. The incidence of postoperative complications in both groups was not significantly different.
Conclusions:
In conclusion our data demonstrate that the automatic cuff pressure and regulation device was useful and reliable in an adult population of intubated patients in the studied pressure range.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessing Blood pressure in the Leg
Preparation: