Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Computational study of the band broadening in two-dimensional etched packed bed columns for on-chip high-performance
P Gzil1, J De Smet, N Vervoort
1Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050 Brussels, Belgium. pgzil@vub.ac.be
Abstract:
The chromatographic performance of several straightforward two-dimensional etched packed bed column lay-outs (equilaterally staggered arrays of, respectively, circular, hexagonal, and diamond-like pillars) has been compared using commercial computational fluid dynamics software. In all cases, the bed porosity was kept at epsilon = 0.4 and a retained component with zone capacity ratio k" = 2 was considered. Exploring the use of six different possible characteristic dimensions to bring the Van Deemter plots of the three different considered particle shapes into agreement, none of them yielded a perfect agreement. Using the pillar volume-based equivalent cylinder diameter (deq) as the characteristic dimension, the diamond-like pillars yielded a significantly smaller h(min) value than the cylinders and the hexagons (h(min) approximately equal to 0.74 for the former versus h(min) approximately equal to 0.83 for the two latter). Including the flow resistance into the analysis, it was found that the "hydrodynamic" shape of the particles has an important influence on the separation impedance E. The more axially elongated diamond pillars yielded an Emin number as small Emin = 180 (for a retained component with k" = 2), i.e. about 40% smaller than the cylinders and the hexagons (Emin = 300-330). The obtained h(min) and Emin values are also significantly smaller than the values often cited for the best possible packed bed HPLC columns. We believe this is a consequence of the assumed perfect homogeneity of the etched structures, and hence hints at the potential benefits of perfectly ordered chromatographic columns, as was already inferred by Knox [J. Chromatogr. A 831 (1999) 3; 960 (2002) 7] and He et al. [Anal. Chem. 70 (1998) 3790].
Related Concept Videos
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...