Related Experiment Video
Updated: Aug 25, 2026

Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
Chromatographic and column stability at pH 7 of a C18 dimethylurea polar stationary phase
César Ricardo Silva1, Carol H Collins, Isabel Cristina Sales Fontes Jardim
1Instituto de Química, Universidade Estadual de Campinas, Caixa Postal 6154, 13083-970 Campinas, São Paulo, Brazil.
Abstract:
Chromatographic evaluations of a C18 dimethylurea phase in 150 mm x 3.9 mm HPLC columns were performed using the Tanaka and Engelhardt test mixtures. The applicability of the new C18 dimethylurea phase was also evaluated with a mixture of some herbicides and their metabolites. An artificial aging procedure was also performed by passing a potassium phosphate mobile phase buffered at pH 7.0 through C18 50 mm x 3.9 mm dimethylurea columns. The column stability was evaluated by means of the chromatographic parameters obtained for the separation of some compounds from the Neue test mixture, using apolar, polar and highly basic analytes.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Ion-Exchange Chromatography
Analyte Adsorption and Distribution
High-Performance Liquid Chromatography: Elution Process