Gated blood-pool SPECT automated versus manual left ventricular function calculations
R H J A Slart1, L Poot, D A Piers
1Department of Nuclear Medicine, University Hospital Groningen, The Netherlands. r.slart@nucl.azg.nl
Gated blood-pool single photon emission computed tomographic (GBPS) left ventricular ejection fraction (LVEF) calculations showed good correlation with planar gated blood-pool imaging (GBPI). However, GBPS values differed significantly from GBPI, indicating they are not equivalent.
Area of Science:
- Nuclear Cardiology
- Cardiovascular Imaging
- Medical Physics
Background:
- Planar gated blood-pool imaging (GBPI) is a standard for assessing left ventricular (LV) function non-invasively.
- Gated blood-pool single photon emission computed tomographic (GBPS) offers simultaneous visualization of all cardiac chambers.
- Evaluating the agreement between automated/manual GBPS and conventional GBPI left ventricular ejection fraction (LVEF) is crucial.
Purpose of the Study:
- To assess the agreement between automated and manual LVEF calculations derived from GBPS.
- To compare these GBPS-derived LVEF values with conventional GBPI LVEF measurements.
Main Methods:
- Performed GBPI in 22 patients, followed by GBPS data acquisition.
- Calculated LVEF using NuSMUGA software (automated and manual methods) from GBPS data.
- Correlated automated and manual GBPS LVEF with conventional GBPI LVEF.
Main Results:
- Automated GBPS LVEF correlated well with GBPI LVEF (r = 0.88).
- Manual GBPS LVEF also correlated well with GBPI LVEF (r = 0.90).
- Both automated and manual GBPS LVEF values showed significant differences compared to GBPI LVEF, and manual GBPS LVEF was significantly higher than automated GBPS LVEF.
Conclusions:
- NuSMUGA GBPS software provides reproducible LVEF values.
- Automated and manual GBPS LVEF correlate well with conventional GBPI.
- GBPS LVEF calculations (both automated and manual) are not equivalent to GBPI LVEF measurements.