Matrix effects in clinical immunoassays and the effect of preheating and cooling analytical samples
Hiroshi Yoshida1, Yuji Imafuku, Toshihiko Nagai
1Department of Clinical Laboratory Medicine, Fukushima Medical University School of Medicine, Fukushima, Japan. hyoshida@fmu.ac.jp
Abstract:
Immunological reactions are influenced by various factors including antigens, antibodies and other variables. We focused on two items: i) matrix effects, especially of detergents and ii) temperature effects: preheating sera, especially effects on rheumatoid factor (RF) measurement and false-positive reactions in ELISAs, and cold storage of sera, especially effects on complement. Among various additives, detergents affected the agglutination reaction for fecal hemoglobin and hepatitis B surface (HBs) antigen. Some of the detergents examined abolished these antigenicities, however, polyethylenglycols enhanced the reactions. Heat-inactivation of sera at 56 degrees C for 30 min was employed in serological testing. However, in RF measurement, 10 min of preheating was sufficient to abolish C1q (subcomponent of C1), which could participate in the agglutination reaction. In ELISA for antibodies, false-positive reactions were caused by preheating sera. By the analyses of assays for antibodies to hepatitis C virus (HCV) and cardiolipin, it was found that they were induced by immunoglobulin G (IgG) modified by preheating. Cold storage induced activation of complement (cold activation) in anti-HCV antibody positive sera. CH50 titers in the sera were lowered by one cycle of freezing at -20 degrees C and thawing, and the decrease was affected by the containers.
More Related Videos
09:08Developing a Salivary Antibody Multiplex Immunoassay to Measure Human Exposure to Environmental Pathogens
Published on: September 12, 2016
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020