Related Experiment Video
Updated: Aug 25, 2026

Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
Published on: December 22, 2023
Determination of organometallics in intra-oral air by LT-GC/ICP-MS
T Riechmann1, A V Hirner, J Feldmann
1Institute of Environmental Analytical Chemistry, University of Essen, D-45117, Essen, Germany.
Abstract:
Low temperature gas chromatography (LT-GC) coupled on-line with inductively coupled plasma mass spectrometry (ICP-MS) has been used to identify volatile metal and metalloid compounds in human breath. After cryogenic sampling, the gas sample has been separated without any clean-up by increasing the temperature (-100 to +200 degrees C). Simultaneous determination of 11 elements with ICP-MS was used for screening analysis. The detection limits of volatile compounds in intra-oral air are in the range of ng m(-3). Dimethyl selenium has been determined in each gas sample from six test persons in the range of 0.08 to 0.98 microg m(-3).
Related Concept Videos
Properties of Organometallic Compounds
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences