Related Experiment Video
Updated: Aug 25, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Monte Carlo dosimetric characterization of the Cs-137 selectron/LDR source: evaluation of applicator attenuation and
J Pérez-Calatayud1, D Granero, F Ballester
1Department of Atomic, Molecular and Nuclear Physics and IFIC, University of Valencia-CSIC, Burjassot, Spain.
Abstract:
The purpose of this study is to calculate the dose rate distribution for the Amersham Cs-137 pellet source used in brachytherapy with the Selectron low-dose-rate remote afterloading system in gynaecological applications using the Monte Carlo code GEANT4. The absolute dose rate distribution for the pellet source was obtained and presented as a one-dimensional absolute dose rate table as well as in the Task Group 43 dose-calculation formalism. In this study, excellent agreement was found between the point source theoretical model using fitted polynomial values and Monte Carlo calculations of the dose rate distribution for the pellet source. A comparison study was also made between the dose rate distribution obtained from a complete Monte Carlo simulation (Cs-137 pellet sources + remote afterloading system plastic guide tube + gynaecological applicator) and that calculated by applying the superposition principle to Monte Carlo data of the individual pellet sources. The data were obtained for a portion of uterine tandem of typical train source configurations. Significant differences with a strong dependence on polar angle have been found that must be kept in mind for clinical dosimetry.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017