Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Characteristics and performance of a micro-MOSFET: an "imageable" dosimeter for image-guided radiotherapy
Carl G Rowbottoma1, David A Jaffray
1Department of Radiation Oncology, William Beaumont Hospital, 3601 W 13 Mile Road, Royal Oak, Michigan 48073, USA. carl.rowbottom@physics.org
Abstract:
The performance and characteristics of a miniature metal oxide semiconductor field effect transistor (micro-MOSFET) detector was investigated for its potential application to integral system tests for image-guided radiotherapy. In particular, the position of peak response to a slit of radiation was determined for the three principal axes to define the co-ordinates for the center of the active volume of the detector. This was compared to the radiographically determined center of the micro-MOSFET visible using cone-beam CT. Additionally, the angular sensitivity of the micro-MOSFET was measured. The micro-MOSFETs are clearly visible on the cone-beam CT images, and produce no artifacts. The center of the active volume of the micro-MOSFET aligned with the center of the visible micro-MOSFET on the cone-beam CT images for the x and y axes to within 0.20 mm and 0.15 mm, respectively. In z, the long axis of the detector, the peak response was found to be 0.79 mm from the tip of the visible micro-MOSFET. Repeat experiments verified that the position of the peak response of the micro-MOSFET was reproducible. The micro-MOSFET response for 360 degrees of rotation in the axial plane to the micro-MOSFET was +/-2%, consistent with values quoted by the manufacturer. The location of the active volume of the micro-MOSFETs under investigation can be determined from the centroid of the visible micro-MOSFET on cone-beam CT images. The CT centroid position corresponds closely to the center of the detector response to radiation. The ability to use the cone-beam CT to locate the active volume to within 0.20 mm allows their use in an integral system test for the imaging of and dose delivery to a phantom containing an array of micro-MOSFETs. The small angular sensitivity allows the investigation of noncoplanar beams.
Related Concept Videos
MOSFET
In an n-MOSFET, the structure includes n-type source and drain...
Characteristics of MOSFET
Various vital parameters influence their functionality, which is crucial for theory and electronics applications. First, channel dimensions, precisely length, and width, are pivotal. The size of these channels affects the transistor's ability to carry current and switching speeds; shorter channels typically enable quicker...
MOSFET Amplifiers
MOSFET: Enhancement Mode
In their basic form, enhancement-mode MOSFETs are typically non-conductive when the gate-source voltage (Vgs) is zero. This default 'off' state means no current...
MOSFET: Depletion Mode
The primary characteristic of depletion-mode MOSFETs is their ability to conduct current between the drain and source terminals without gate bias. This inherent conductivity arises...