Related Experiment Video
Updated: Aug 24, 2026

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose
Published on: December 30, 2025
New glucose biosensor based on glucose oxidase-immobilized gelatin film coated electrodes
Sibel Sungur1, Emel Emregül, Gülendem Günendi
1Faculty of Science, Department of Chemistry, Biochemistry, Ankara University, 06100 Beşelver, Ankara, Turkey. sungur@science.ankara.edu.tr
Abstract:
In this article we describe the preparation of a new glucose biosensor based on electrode coating with glucose oxidase-immobilized gelatin. Glucose oxidase was immobilized onto gelatin by cross-linking with chromium III acetate. Properties of the immobilized glucose oxidase enzyme electrode have been studied and results were compared with free enzyme electrode. Effect of temperature and pH variations, kinetic properties, and re-usability were among the subjects analyzed. Also interfering effect of the electro-active compounds was investigated and their influence on the behavior of glucose oxidase electrode was determined. The response time of the developed glucose oxidase enzyme electrode was 60 s with a detection limit of 0.25 mM and preserved its accuracy within the 2 months period of reuse experiments.
Related Concept Videos
Amperometry: Overview
Microbial Biosensors