Cellular phone interference with the operation of mechanical ventilators
Cheryl I Shaw1, Robert M Kacmarek, Rickey L Hampton
1Department of Biomedical Engineering, Massachusetts General Hospital, Boston, MA, USA.
Objective:
To determine whether a cellular phone would interfere with the operation of mechanical ventilators.
Design:
Laboratory study.
Setting:
University medical center.
Subjects:
Fourteen mechanical ventilators.
Interventions:
We evaluated change in operation and malfunction of the mechanical ventilators.
Measurements And Main Results:
The cellular phone (Nokia 6120i) was computer controlled, operating at 828.750 MHz analog modulation. It was operated at 16, 40, 100, 250, and 600 mW, 30 cm from the floor and 30, 15, and <3 cm from all sides of each ventilator. Six of the 14 ventilators tested malfunctioned when a cellular phone at maximum power output was placed < or =15 cm from the device. None of these responses were considered immediately life threatening except for the response of the Puritan Bennett 840, which stopped ventilating when the cellular phone at maximum power output was placed < or =30 cm from the ventilator. One ventilator doubled the ventilatory rate and another increased the displayed tidal volume from 350 to 1033 mL. In one of the infant ventilators, displayed tidal volume increased from 21 to 100 mL. In another ventilator, the high respiratory rate alarm sounded but the rate had not changed.
Conclusions:
In a controlled laboratory setting, cellular phones placed in close proximity to some commercially available intensive care ventilators can cause malfunctions, including irrecoverable cessation of ventilation. This is most likely to occur if the cellular phone is <30 cm from the device and ringing. Based on our data and the available literature, we believe it is reasonably safe to permit the use of cellular phones in the intensive care unit, as long as they are kept > or =3 feet from all medical devices. The current electromagnetic compatibility standards for mechanical ventilators are inadequate to prevent malfunction. Manufacturers should ensure that their products are not affected by wireless technology even when placed immediately next to the device.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...