Orthotripsy mimicking asystole.

Stephen C Vlay1

  • 1Stony Brook Arrhythmia Study and Sudden Death Prevention Center, Division of Cardiology, Department of Medicine, Stony Brook University, New York, USA. svlay@notes.cc.sunysb.edu

Pacing and Clinical Electrophysiology : PACE
|April 14, 2004
PubMed
Summary

An implantable loop recorder (ILR) can detect apparent arrhythmias during procedures like orthotripsy. However, electrostatic discharge can cause artifact, mimicking serious cardiac events like ventricular asystole.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Cold case files: A tale of three sisters - solved by genetic testing with implications for LMNA cardiomyopathy.

Pacing and clinical electrophysiology : PACE·2024
Same author

Thirty-five-year follow-up of 3 generations of a family with LMNA cardiomyopathy.

Heart rhythm·2024
Same author

Unregulated online sales of cardiac implantable electronic devices in the United States: A six-month assessment.

Heart rhythm O2·2021
Same author

Global disparities in cardiac pacemaker therapy: Problem statement, potential solution, and call to action.

Heart rhythm·2018
Same author

Worldwide pacemaker and defibrillator reuse: Systematic review and meta-analysis of contemporary trials.

Pacing and clinical electrophysiology : PACE·2018
Same author

Barriers to pacemaker and ICD recycling.

Pacing and clinical electrophysiology : PACE·2018

Area of Science:

  • Cardiology
  • Medical Devices
  • Electrophysiology

Background:

  • Implantable loop recorders (ILRs) are used for long-term cardiac monitoring.
  • Orthotripsy procedures can potentially interfere with cardiac monitoring devices.

Purpose of the Study:

  • To evaluate an apparent arrhythmia detected by an ILR during an orthotripsy procedure.
  • To differentiate true arrhythmias from device-related artifacts.

Main Methods:

  • An ILR was implanted to investigate dizziness.
  • The ILR recorded an event during an orthotripsy procedure.
  • Cardiac tracing analysis was performed to identify the event's origin.

Main Results:

  • An apparent arrhythmia was documented by the ILR during orthotripsy.
  • Review of the tracing identified the event as artifact.
  • The artifact was attributed to electrostatic discharge.

Conclusions:

  • Electrostatic discharge during orthotripsy can create artifactual signals.
  • These artifacts may mimic serious arrhythmias such as ventricular asystole.
  • Careful review of ILR tracings is crucial to avoid misinterpretation of procedural artifacts.